Piezo-Metrics' Model MP40 is a flush mountable, miniature, low cost, general use pressure transducers. These transducers are made of corrosive resistant titanium (6AL4V). They utilize high reliability bulk type semiconductor strain gages bonded directly to the inside surface of the machined diaphragm.
Titanium hermetic headers permit offering the MP40 series with absolute or sealed reference options. The hermetic header is located just behind the strain gage diaphragm section of the sensor. Sealing the sensitive gages to an inert environment dramatically increases stability and reliability. Long term stability versions are available. High vibration options are also available.
A |
B |
C |
D |
Units |
|
---|---|---|---|---|---|
Balance (Zero) | ±3.00 | ±2.00 | ±2.00 | ±2.00 mV | mV |
Full Scale Sensitivity (MP40) | 2.00±2.00 | 2.00±2.00 | 20.0±2.00 | 20.0±2.00 | mV/V |
Static Error Band | ±0.50 | ±.25 | ±0.25 | ±0.25 | %FS*BFSL |
Thermal Balance (Zero) Shift | ±0.02 | ±0.02 | ±0.01 | ±0.05 | %FS/°F |
Thermal Sensitivity Shift | ±0.02 | ±0.02 | ±0.02 | ±0.025 | %FS/°F |
Standard Long Term Stability | >± 2.00 %FS/Year Max. | ||||
Long Term Stability Options | L10= <1.0 | L5=<0.5 | L2=<0.2 | L1=<0.1 | %FS/°F |
Standard Pressure Ranges | 300, 500, 1K, 2K, 3K, 5K, 7K psi |
Non Standard Pressure Ranges | 30, 35, 40, ,50 100, 150, 200, 250, 10K psi |
Other Ranges Possible | Consult Factor |
Over Pressure (No change out of spec.) | 2.0x Range |
Usable Pressure (Small change in balance) | 3.0x Range |
Burst Pressure | 6.0x Range |
Storage and Operating Temperature | 0° to 180°F |
Compensated Temperature Range | 30° to 130°F |
Other Temperature Compensated Ranges | Consult Factor |
Acceleration (Consult factory for high vibration options) | 100g’s any axis |
Media Compatibility (wetted surface) | Liquids & Gases compatible with Titanium 6AL4V |
Input | 5.0 Volts DC or AC |
Maximum Voltage | 30 V for short periods |
Input Resistance (Nominal) | 1400 400 Ohms |
Output Resistance (Nominal) | 850 200 Ohms |
Insulation Resistance Minimum | 50 MOhms @ 50 VDC |